- Contact Person : Ms. Dai Sophie
- Company Name : Taixing Ruisheng Co., Ltd.
- Tel : 86-523-87690688
- Fax : 86-523-82919867
- Address : Jiangsu,Taixing,No.39,Liberation Bridge,Taixing,Jiangsu,225400
- Country/Region : China
- Zip : 225400
High Frequency Mobile C-arm Machine
High Frequency Mobile C-arm Machine
Specifications
Fluoroscopic Capacity
Max rated capacity : Tube Current 4mA, Tube Voltage 110kV
Automatic Fluoroscopy : Tube Voltage:40kV~110kV adjust
automatically
Tube Current : 0.3mA~4mA set manually
Manual Fluoroscopy : Continuous Tube Voltage:40kV~110kV
Continuous Tube Current : 0.3mA~4mA
Pulse Fluoroscopy : Continuous Tube Voltage 40kV~110kV
Continuous Tube Current 4.1mA~8mA
Photography Capacity
Max rated capacity : 3.5 KW
Tube Voltage and Cur-
rent Combination : 40kV~49kV 1 ~125 mAs
50kV~59kV 1 ~110 mAs
60kV~69kV 1 ~90 mAs
70kV~79kV 1 ~80 mAs
80kV~89kV 1 ~71 mAs
90kV~99kV 1 ~63 mAs
100kV~110kV 1 ~40 mAs
Plateholder Size : 200mm×250mm(8″×10″)or
250mm×300mm(10″×12″)
X-ray Tube
X-ray Tube Special for High Frequency
Fixed anode Dual-focus : large focus: 1.5, small focus: 0.6
Inverter Frequency: 40KHz
Thermal capacity: 35KJ (47KHU)
VideoSystem
Image Intensifier : Image Intensifier made by TOSHIBA (9″)
CCD Vidicon : Ultra low-light CCD camera imported from Japan
Monitor : Horizontal 1000 lines and vertical 800 lines,
Bandwidth : 12.5MHz, Image/sec : 75
CCU(central control) : Recursive Filter: K=8, 8 images storage,
image upright, image overturn,positive &
negative image LIH(last image freeze, and
OSD(monitor display)
Structure :
Directive wheel : ±90°revolution,can freely change the moving
direction of the unit.
Ascending & Descend-
ing Range of Pillar : ≥400mm
C-arm : Forward and Backward Movement: 200mm
Revolution around Horizontal Axis: ±180°
Revolution around Vertical Axis : ±15°
Slip on orbit: 120°(+90°~ -30°)
Packing Details : 2440mm×1100mm×1420mm
N.W. 300kgs G.W.400kgs
Features
1,With a compact appearance, and easy to operate;
2,Clinical Video system with high performance,8 images storage
volume, and one 14''high-resolution monitors;
3,Automatically track fluoroscopy to make the image brightness and
clearness optimum;
4,Automatically store the last image of fluoroscopy to advantage
diagnosis;
5,With a high-quality knockdown X-ray generator to reduce radiation
High Frequency Mobile C-arm Machine